Multisource cone beam computed tomography using a carbon nanotube x-ray source array

MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1(2024)

引用 0|浏览8
暂无评分
摘要
The purpose of this study is to develop and evaluate a functional multisource cone beam computed tomography (ms-CBCT) scanner to mitigate some of the main limitations of the current CBCT. The benchtop ms-CBCT utilizes a carbon nanotube (CNT) field emission source array to generate multiple narrowly collimated and rapidly scanning x-ray beams, each illuminating a section of the object and collectively covering the region of interest. A contrast phantom and a Defrise phantom were imaged by the ms-CBCT, the ms-CBCT operating in the convetional CBCT configuration, and a clinical CBCT. The results show the ms-CBCT reduces the spatial nonuniformity and root-mean-square error (RMSE) of the CT HU values by respectively 75% and 60%, essentially eliminates the cone beam artifacts, increases the effective axial coverage, and improves the CNR by 30% similar to 50% compared to the conventional CBCT at a comparable imaging dose. The results show that the ms-CBCT can potentially provides the performance of an MDCT while maintaining the essential attributes of a CBCT including volumetric imaging, low dose, affordability, and compact design.
更多
查看译文
关键词
Multisource CT,CBCT,CNT X-ray,HU uniformity,HU accuracy,cone beam artifacts