Outcomes After Heartmate 3 Left Ventricular Assist Device Implantation Using a 10 mm Outflow Graft.

ASAIO journal (American Society for Artificial Internal Organs : 1992)(2024)

引用 0|浏览12
暂无评分
摘要
The presence of adhesions and patent bypass grafts may create challenges for standard 14 mm outflow graft placement during left ventricular assist device implantation. We retrospectively describe our experience using a 10 mm Bioline Fusion graft (Getinge, Goteborg, Sweden) as the outflow graft in patients undergoing primary Heartmate 3 (Abbott, Abbott Park, IL) implantation. One hundred one patients underwent Heartmate 3 left ventricular assist device implantation, 80% via a thoracotomy approach, with the standard 14 mm outflow graft (78) or a 10 mm Bioline Fusion outflow graft (23). Initial postoperative rotor speed-to-flow ratio (the revolutions per minutes (RPMs) required to achieve a given flow) was significantly higher in 10 mm graft patients (1,472 vs. 1,283 RPM/L/min; p = 0.03), suggesting elevated resistance in the smaller graft. Furthermore, the initial postoperative vasoactive-inotrope score was higher in the 10 mm graft patients (24.1 vs. 17.6; p = 0.022). Postoperative outcomes were similar between groups. In conclusion, the use of a 10 mm graft was associated with higher RPMs needed to generate a given flow and a higher vasoactive-inotrope score, but these differences were not associated with increased right ventricular failure or mortality.