Location and Length of Gap at the Interface of an Idealized Cemented Hip Implant Subjected to Bending

Engineering in Medicine and Biology Society, 2000 Proceedings of the 22nd Annual International Conference of the IEEE(2000)

引用 0|浏览1
暂无评分
摘要
The uncertainties of the residual stresses developed during cement polymerization and of the coefficient of friction were considered in the 3-D finite element model using nonlinear contact elements to model the debonded interface. An accurate modeling of the residual stresses and coefficient of friction are important to predict the gap lengths; for an interference of 0.05 mm and a coefficient of friction of 0.4, no gap was observed. These indicate possible true gap locations at the interface which could lead to initial loss of fixation of the implant
更多
查看译文
关键词
bending,biomechanics,finite element analysis,internal stresses,mechanical contact,orthopaedics,physiological models,prosthetics,3-D finite element model,bending,cement polymerization,coefficient of friction,debonded interface,gap length,gap location,idealized cemented hip implant,initial loss of fixation,nonlinear contact elements,residual stresses