A Rare Complication of Tension Band Fixation of Olecranon Osteotomy: Distal Migration of K-wire.

JOINT DISEASES AND RELATED SURGERY(2024)

引用 0|浏览3
暂无评分
摘要
Tension band wiring (TBW) is one of the most commonly used fixation techniques to fix olecranon osteotomies. Hardware prominence has been the most commonly reported complication of TBW. However, distal migration of Kirschner (K)-wire after TBW fixation for olecranon osteotomy has not been reported. In this case report, we presented distal migration of K-wire detected nine months after initial surgery in a 46-year-old male patient. The patient was operated on for an intraarticular distal humerus fracture using an olecranon osteotomy. The osteotomy was fixed with TBW fixation. The patient missed routine follow-ups and presented to the outpatient clinic with a complaint of skin irritation at the elbow nine months after the surgery. On radiological examination, distal migration of one K-wire was detected. The K-wire was surgically removed without any complication. Physicians should be aware of possible complications of TBW and remove fixation after fracture union to avoid unexpected complications.
更多
查看译文
关键词
Complication,distal humerus,elbow,pin migration,trauma