Rapid calibration curve generation using single-sheet radiochromic film for proton-based stereotactic radiosurgery verification

Radiation Physics and Chemistry(2024)

引用 0|浏览3
暂无评分
摘要
Proton-based stereotactic radiosurgery (PSRS) holds promise for treating benign tumours, with patient-specific quality assurance (PSQA) playing a crucial role in ensuring high-quality clinical services. EBT3 films are a perfect tool for PSQA, particularly owing to their compatibility with small-field dosimetry. However, a successful absolute EBT3 dosimetry relies on detailed dose calibration, consuming much time and films. A novel approach has been developed in this study and tested to establish a calibration curve using a single sheet of EBT3 radiochromic films exposed to six multi-leaf collimator (MLC) fields. Notably, this method allows for the completion of a calibration curve in just 15 min, enabling the practical generation of new EBT3 calibrations daily. This single-sheet approach can generate a dose calibration curve with an error of less than 1.5% across the entire dose range and is well agreed to by traditional multi-film calibrations involving large fields, with a discrepancy of less than 0.8%.
更多
查看译文
关键词
Small field,Proton therapy,Proton-based stereotactic radiosurgery (PSRS),EBT3 calibration
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要