Abstract
We characterized the on-board megavoltage imager (MVI) of a magnetic resonance-guided radiotherapy machine for beam output checks. Linearity and repeatability of its dose response were investigated. Alignment relative to the beam under clinical circumstances was evaluated for a year using daily measurements. Linearity and short-term repeatability were excellent. Long-term repeatability drifted 0.8 % per year, which can be overcome by monthly cross calibrations. Long-term alignment was stable. Thus, the MVI has suitable characteristics for beam output checks.
More
Translated text
Key words
Radiotherapy,Quality assurance,MRI-Linac,Unity,Dosimetry,MV imager