A High-Performance System for Weak ECG Real-Time Detection

Sensors(2024)

引用 0|浏览21
暂无评分
摘要
Wearable devices have been widely used for the home monitoring of physical activities and healthcare conditions, among which ambulatory electrocardiogram (ECG) stands out for the diagnostic cardiovascular information it contains. Continuous and unobtrusive sensing often requires the integration of wearable sensors to existing devices such as watches, armband, headphones, etc.; nonetheless, it is difficult to detect high-quality ECG due to the nature of low signal amplitude at these areas. In this paper, a high-performance system with multi-channel signal superposition for weak ECG real-time detection is proposed. Firstly, theoretical analysis and simulation is performed to demonstrate the effectiveness of this system design. The detection system, including electrode array, acquisition board, and the application (APP), is then developed and the electrical characteristics are measured. A common mode rejection ratio (CMRR) of up to 100 dB and input inferred voltage noise below 1 μV are realized. Finally, the technique is implemented in form of ear-worn and armband devices, achieving an SNR over 20 dB. Results are also compared with the simultaneous recording of standard lead I ECG. The correlation between the heart rates derived from experimental and standard signals is higher than 0.99, showing the feasibility of the proposed technique.
更多
查看译文
关键词
ECG,noise suppression,multi-channel system,electrode array