Non-Rigid Motion Compensation for Breast CT

2022 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)(2022)

引用 0|浏览1
暂无评分
摘要
The image quality in dynamic contrast-enhanced breast CT is expected to suffer from motion artifacts due to the extended acquisition time involved. We propose an iterative method for compensation of motion artifacts due to non-rigid movement of the breast. The motion vector field is approximated using b-splines on a sparse grid and its values are found by minimizing errors in the projection domain. We evaluated the method on an anthropomorphic phantom with realistic motion and visual assessment yielded a clear reduction in motion artifacts. Quantitatively, we observed an increase of the structural similarity from 0.9988 to 0.9995 and a decrease of the normalized root mean squared error from 0.1448 to 0.0932.
更多
查看译文
关键词
Motion Compensation,Non-rigid Motion,Breast CT,Root Mean Square Error,Mean Square Error,Motion Artifacts,Dynamic Contrast-enhanced Magnetic Resonance Imaging,Motion Vector,Real Motion,Normalized Root Mean,Anthropomorphic Phantom,Sparse Grid,Projection Domain,Transformer,Gradient Descent,Image Domain,Motion Estimation,Ray Tracing,Motion Field,Forward Projection