谷歌浏览器插件
订阅小程序
在清言上使用

Use of a Rhodamine-Based Chelator in a Microfluidic Paper-Based Analytical Device for the In-Situ Copper Quantification in Natural Waters

Talanta(2024)

引用 0|浏览8
暂无评分
摘要
This work describes the development of a microfluidic paper-based analytical device (mu PAD) for the determination of copper in fresh and marine waters. A functionalized rhodamine-based chelator was synthesized and used as a chromogenic reagent, forming a highly intense pink complex with the analyte. The aim was to create a paper device that offers optimal performance and provides in-situ, rapid and cost-effective analysis in line with World Health Organization guidelines. The influence on the determination of several physical and chemical parameters was evaluated aiming to achieve the best performance. Under optimised conditions, a linear correlation was established in the range of 0.05-0.50 mg L-1 of copper, with a limit of detection of 10 mu g L-1. The accuracy of the proposed method was assessed by comparing the results obtained with the developed mu PAD and the results obtained with Inductively Coupled Plasma measurements (RE < 10 %). Recovery studies were also performed using different types of water samples with no need for any prior sample pre-treatment: tap, well, river and seawater. The average recovery percentage of 101 % (RSD = 4.3 %) was obtained, a clear indication of no multiplicative matrix interferences.
更多
查看译文
关键词
Copper(II) determination,Functionalized copper ligand,In -situ analysis,Natural waters,Paper sensor device