A fully-flexible and thermally adjustable implantable neural probe with a U-turn polyester microchannel

2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC(2023)

Cited 0|Views3
No score
Abstract
This work presents a fully flexible implantable neural probe fabricated with Polydimethylsiloxane (PDMS) and including a thermally-tunable stiffness microchannel filled with Polyester. The probe includes an optimized microfluidics mixer for drug delivery. Polyester, which is solid at room temperature and has a low melting point close to body temperature, is used to decrease the stiffness of the probe after insertion, after getting in contact with tissues. We designed a U-turn microchannel inside the PDMS neural probe and filled it up with melted polyester. The microchannel has a cross-section of 30 mu m x 5 mu m and a length of 14.7 mm. The following probe dimensions were chosen after extensive simulation: thickness = 20 mu m, width = 300 mu m, and length = 7 mm. These values yield a buckling force above 1 mN, which is sufficient for proper insertion into the brain tissues. Simulation results show that the microfluidics mixer with a cross-section of 90 mu m x 5 mu m and a length of 7 mm has optimum performance for the desired flow rate and quantity of drug to deliver. The pressure drop inside the microfluidic channel is less than 0.43 kPa, which is appropriate for PDMS-PDMS bonding, whereas the Reynolds number is near 1.91k in the laminar regime. No leakage or bubble occurred during the experimental validation, which suggests an appropriate pressure and a laminar flow in the channel.
More
Translated text
Key words
Neural probe,Polyester,Microchannel,Stiffness tuning,Microfluidics
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined