Presence of Metallic Components in Transdermal Drug Delivery Systems and Risk of Skin Burns.

FARMACIA HOSPITALARIA(2024)

引用 0|浏览4
暂无评分
摘要
Objective: Skin burns are associated with the presence of metallic components in transdermal drug delivery systems during Magnetic Resonance Imaging, cardioversion, or defibrillation procedures. The aim of the study was to review the presence of metallic components in marketed products of transdermal drug delivery systems in Spain. Method: For each pharmaceutical form, the summary of product characteristics was reviewed. If the information was not provided, manufacturers were contacted. Results: We identified 59 marketed products of transdermal drug delivery systems of 12 different active substances. 59.3% of patches contained metallic components or their presence could not be ruled out. Information regarding the need to remove the patch was only included in 8 summaries of product characteristics (13.6%) A table was elaborated and included the following aspects: product, active substance, manufacturer, need to remove the patch before the exposure to magnetic or electric fields, and references. Conclusion: More than a half of the patches at the time of the study contained metals or their absence could not be confirmed by the manufacturer. However, this information was only included in 13.6% of summaries of product characteristics. (c) 2023 Sociedad Espa & ntilde;ola de Farmacia Hospitalaria (S.E.F.H). Published by Elsevier Espa & ntilde;a, S.L.U. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
更多
查看译文
关键词
Medication safety,Patient safety,Quality improvement,Safety culture,Transdermal drug delivery systems
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要