Thrombectomy aspiration post-market study in acute stroke with the Q aspiration catheter: the TAPAS study

Journal of NeuroInterventional Surgery(2022)

引用 4|浏览5
暂无评分
摘要
Background The Q Aspiration Catheter (MIVI Neuro) has demonstrated greater aspiration flow rates and ingestion forces compared with conventional catheters in vitro. The safety and performance of the Q Catheter was assessed using a direct aspiration first pass technique in patients with acute ischemic stroke at four neurointerventional centers in Spain. Methods We included adult patients who underwent mechanical thrombectomy between March 2019 and March 2020 using the Q Catheter as first-line therapy. Performance endpoints included final successful revascularization of the target vessel (defined as modified thrombolysis in cerebral infarction (mTICI) grade 2B/3), first pass revascularization, and overall Q Catheter revascularization. Safety endpoints were symptomatic intracranial hemorrhage (sICH), embolization to new territory (ENT), and procedural complications. Modified Rankin Scale (mRS) score and all-cause mortality were also assessed. Results Forty-five subjects were enrolled. The Q Catheter successfully navigated to the lesion in 95.5% (43/45) of patients. Final successful mTICI 2B/3 revascularization was achieved in 93.3% (42/45), first pass mTICI 2B/3 revascularization with the Q Catheter was 55.3% (21/38), and overall with Q Catheter mTICI 2B/3 revascularization was 65.8% (25/38). Favorable clinical outcome of mRS 0–2 was achieved in 55.6% (25/45). There were no cases of ENT. sICH and mortality rates were 2.2% (1/45) and 13.3% (6/45), respectively. Conclusion In this multicenter, observational study, the Q Aspiration Catheter used as first-line therapy demonstrated a good and safe profile in terms of navigation, revascularization, and safety in patients with acute ischemic stroke.
更多
查看译文
关键词
thrombectomy aspiration,acute stroke,catheter,post-market
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要