谷歌浏览器插件
订阅小程序
在清言上使用

The Effect of Voxel and Field of View Size on the Volumetric Alteration Artifact of High-Density Materials with 2 Cone Beam Computed Tomography Devices.

Oral surgery, oral medicine, oral pathology and oral radiology(2024)

引用 0|浏览3
暂无评分
摘要
OBJECTIVE:We investigated the influence of voxel and field of view (FOV) sizes on expression of the volumetric alteration artifact (VAA) of 5 high-density materials in 2 cone beam computed tomography (CBCT) devices.STUDY DESIGN:Cylinders of amalgam, cobalt-chromium, gutta-percha, titanium, and zirconium were individually positioned in a polymethyl methacrylate phantom. OP300 Maxio and ProMax 3D Classic CBCT devices were used to acquire images with varying voxel and FOV sizes, totaling 585 scans. Two evaluators segmented the high-density cylinder images to obtain the tomographic volumes. The difference between the tomographic and physical volume of each cylinder (i.e., volumetric alteration) was calculated. Statistical analysis was conducted with multiway ANOVA and the Tukey post hoc test (α = 5%). Evaluator reliability was measured with the intraclass correlation coefficient (ICC).RESULTS:All studied parameters and nearly all interactions influenced the VAA (P < .05). The post hoc test demonstrated less volumetric alteration for the smallest voxel sizes, 61 × 78 mm FOV, and gutta-percha for OP300, and for the smallest voxel sizes, 80 × 80 mm FOV, and gutta-percha and titanium for ProMax 3D (P < .05). The ICC demonstrated perfect reliability (1.00).CONCLUSIONS:Voxel and FOV sizes influenced VAA expression. Using smaller voxel sizes, the 61 × 78 mm FOV for OP300 and the 80 × 80 mm FOV for ProMax 3D, and materials with lower density and lower atomic number reduced VAA expression.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要