Parylene C is well-known as an encapsulation material for medical implants. Within the approach of miniaturization and automatization of a bone distractor, piezoelectric actuators were encapsulated with Parylene C. The stretchability of the polymer was investigated with respect to the encapsulation functionality of piezoelectric chips. We determined a linear yield strain of 1% of approximately 12-mu m-thick Parylene C foil. Parylene C encapsulation withstands the mechanical stress of a minimum of 5x105 duty cycles by continuous actuation. The experiments demonstrate that elongation of the encapsulation on piezoelectric actuators and thus the elongation of Parylene C up to 0.8 mm are feasible.
更多
查看译文
关键词
Parylene C,encapsulation,piezoelectric actuator,stretchability,active implant,polymer