Hearing rehabilitation through middle ear surgery can be challenging, particularly for patients with chronic otitis media, but new hearing devices offer opportunity to reach this goal. The aim was to compare hearing outcomes and quality of life in patients who were converted from a Baha Attract system to the Osia system. Prospective observational study, in which each subject acted as their own control. Six patients who were converted from a Baha Attract to an Osia system performed audiometric and quality of life assessments with different sound processor at fitting, 3, 6 and 12 months. The speech performance in noise with the Osia system at 12 months, expressed as dB SNR, was compared to the baseline condition (Baha 5). The PTA4 hearing thresholds improved from 42.6 ± 11 with Baha 5 Power to 34.8 ± 13.3 with Baha 5 SuperPower and to 25.4 ± 3.5 with the Osia at 12 months, leading to a significant functional gain of 17.2 ± 10.9 dB vs Baha 5 (p < 0.02). Speech understanding in both quiet and noise was clinically improved reaching a mean SNR of less than 1 dB at 12 months with the Osia system. Quality of life outcomes improved by more than 20