CT Annulus Sizing Prior to Transcatheter Aortic Valve Replacement (TAVR): Evaluation of Free-Breathing Versus Breath-Holding Acquisition
EUROPEAN RADIOLOGY(2023)
Abstract
To assess feasibility and accuracy of aortic annulus measurements using cardiac computed tomography angiography (CTA) performed during free-breathing prior to transcatheter aortic valve replacement (TAVR). Sixty consecutive TAVR candidates underwent free-breathing wide-detector cardiac CTA, followed by a percutaneous valve replacement. For each, a theoretical valve size was suggested through CT measurements of the annulus, then compared to the size of the actual implanted transcatheter heart valve (THV). The procedural success and the 30-day outcomes were collected. Image quality of the annulus was also studied according to subjective and objective criteria. Data of a control group of 60 patients previously evaluated on breath-holding were also evaluated. A total of 120 patients (mean age, 83 years ± 7, 60 men) were evaluated. All CT acquisitions provided sufficient image quality allowing precise annulus measurements. Mean attenuation (p < 0.001) and image noise (p = 0.01) were higher in the free-breathing group, while image quality was comparable (p = 0.36). The agreement rate between CT-suggested valve size and THV implanted size was comparable, estimated at 87
MoreTranslated text
Key words
Breath-holding,Transcatheter aortic valve replacement,Multidetector computed tomography,Aortic valve stenosis