Ophthalmic fundus camera design based on freeform surface for reducing refractive error sensitivity
Optics and Lasers in Engineering(2023)
摘要
The eye has a remarkably complex structure and biomechanics. An imbalance between the forces acting on several muscles and the tissue properties may alter or even preclude vision. With the widespread use of electronics, refractive errors have become a common problem, making clear fundus imaging difficult. To overcome this challenge, freeform surfaces have emerged as a potential solution, enabling compact, low-cost, and user-friendly systems that are insensitive to refractive errors. We propose a fundus camera based on a freeform surface that can image the fundus without the influence of refractive errors. A special front lens was designed to image the pupil on a freeform lens that can realize phase modulation. The system performs well providing a field of view of 40∘ and a pupil diameter of 3 mm for refractive errors ranging from -5D to +5D. The volume of the system was less than 35 mm × 35 mm × 135 mm.
更多查看译文
关键词
ophthalmic fundus camera design,freeform surface
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要