Design of a Compact Long Counter with an Improved Response Using Multiple Point-Like Thermal Neutron Counters

Pengxiang Wang,Zhimeng Hu, Xucheng Ma,Pin Gong,Guoqiang Zhong,Giuseppe Gorini, Baolong Chen,Xiaobin Tang

˜The œEuropean physical journal plus(2023)

引用 0|浏览14
暂无评分
摘要
A long counter with multiple point-like thermal neutron counters positioned in a cylindrical neutron moderator was investigated to improve the performance of the long counter, including flatter fluence response in a wider neutron energy range and also the compactness. The summation of the weighted counts of each thermal neutron counter yields the response of the long counter. In this study, silicon carbide (SiC) coated with 6LiF convertor was used as the thermal neutron counter. In a further effort to balance the performance and structural complexity of long counters with multiple thermal neutron counters, the number and positions of these counters were first studied and optimized. As a result, the long counter with four specially positioned SiC counters showed the best performance. Besides, we also implemented a copper annular into the moderator to compensate for the long counter’s response above a few MeV. The response function was relatively flat in the energy range from 10 eV to 25 MeV, with a relative standard deviation of approximately 1.78
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要