CG02N Analyzer Accurately Measures Fibrinogen Level in Whole Blood in the Presence of Low- or High-Dose Heparin
INDIAN JOURNAL OF HEMATOLOGY AND BLOOD TRANSFUSION(2024)
摘要
Rapid fibrinogen (Fbg) evaluation is important in patients with massive bleeding during severe trauma and those undergoing major surgery. However, there are only a few studies on the point-of-care Fbg analyzer. In this study, we aimed to investigate the accuracy of Fbg level measured using CG02N, with whole blood contained in lithium-heparinized syringes with two different concentrations of heparin. Blood samples were collected in lithium-heparinized tubes, namely PREZA-PAK (R) II (low-dose heparin group [LG], 7 IU/mL) and Pro-Vent (R) Plus (high-dose heparin group [HG], 70.5 IU/mL). The Fbg levels in LG and HG were compared with those of citrated plasma Fbg (standard-Fbg). Strong correlations with respect to the Fbg level were observed between standard-Fbg and LG or HG (r = 0.968, p < 0.0001; r = 0.970, p < 0.0001, respectively). We demonstrated that the Fbg level in whole-blood samples was accurately assessed by CG02N and not affected by low- or high-dose heparin.
更多查看译文
关键词
Trauma,Fibrinogen,Point of care testing,Massive bleeding,Heparinized tube