Clinical Applicability of Automated Hematology Analyzer Research Parameters for Disease Diagnosis: a Reference Interval Study.

Clinical laboratory(2023)

引用 0|浏览11
暂无评分
摘要
BACKGROUND:We established reference intervals for research parameters of complete blood cell count and examined their usefulness for diagnosing certain diseases.METHODS:Reference intervals for 26 basic and 38 research parameters were established for 3,457 and 1,325 men and 2,742 and 830 women aged 20 - 59 and ≥ 60 years, respectively. Research parameter values for patients with iron deficiency anemia (IDA), appendicitis, sepsis, and myelodysplastic syndromes (MDS) were compared against gender- and age-matched reference values.RESULTS:Seven basic and 10 research parameters among men and one research parameter among women required partitioning by age. No partitioning by gender was required. Further, 67% patients with IDA showed micro red blood cell ratio values above the upper reference limits of their corresponding age and gender subgroups; 3% and 5% patients with appendicitis showed immature granulocyte percentages and counts above the upper reference limits, respectively; 12% - 42% of patients with sepsis showed numerous values exceeded their reference limits, and 67% and 100% patients with MDS showed neutrophil cell complexity and structural dispersion values outside their reference ranges, respectively.CONCLUSIONS:Overall, < 60% of research parameter values were outside their reference ranges among most patients, indicating their limited diagnostic usefulness.