The Hidden Cochlear Implant.

Journal of laryngology and otology/˜The œjournal of laryngology & otology(2023)

引用 0|浏览18
暂无评分
摘要
OBJECTIVE:The hidden cochlear implant concept has two data transmission methods: Bluetooth low energy and transtympanic optical data transfer systems. This study aimed to present the hidden cochlear implant and compare the test results with the existing fully implanted cochlear implant.METHOD:The Bluetooth low energy module was implanted into the implant bed. For the transtympanic optical data transfer tests, a receiver was passed through the posterior tympanotomy, and the transmitter was placed in the ear canal.RESULTS:The Bluetooth low energy module range was 5.2-17.5 m. Transtympanic optical data transfer reached a rate of 1 Mbit/s and had 99.22 per cent accuracy. Despite various obstacles, the accuracy of the transtympanic optical data transfer was more than 99 per cent with a 250 Kbit/s rate. The average power consumption was 310 mW for the implanted Bluetooth low energy module and 41 mW for the transtympanic optical data transfer receiver.CONCLUSION:Bluetooth low energy is suitable to be used transcutaneously. Transtympanic optical data transfer is an effective and promising technology. Hidden use cochlear implants aim to have the aesthetics of a fully implantable cochlear implant with higher reliability and a magnet-free design with smart device integration.
更多
查看译文
关键词
Cochlear Implants,Inner Ear,Quality Of Life