In Silico Prediction of Thrombosis Risk in a Ventricular Assist Device Supported Right Heart: the Impact of Cannulation Site

ASAIO journal(2022)

引用 4|浏览0
暂无评分
摘要
Right ventricular assist device (RVAD) associated thrombosis is a serious complication that may arise due to unfavorable blood flow dynamics (blood stasis) caused by RVAD cannula protrusion within the chambers. This study aims to investigate the thrombosis risk of cannulation via the right atrium (RA) and right ventricle (RV) (diaphragmatic) under full RVAD support using computational fluid dynamics. A HeartWare HVAD inflow cannula was virtually implanted in either the RA or RV of a rigid-walled right heart geometry (including RA, RV, superior, and inferior vena cava) extracted from computed tomography data of a biventricular support patient. Transient simulations, validated with particle image velocimetry, were performed with constant inflow. Thrombosis risk was predicted by analyzing the time-averaged blood velocity, blood stagnation volume, washout rate, and blood residence time (BRT). Results showed that RA cannulation disturbed the physiological swirling flow structure which can be found in an uncannulated RA. This led to a large low-velocity recirculation flow in the RV, increasing the thrombosis risk. Contrarily, RV diaphragmatic cannulation showed better preservation of swirling flow in the RA and flow ejection into the RV. Consequently, RV diaphragmatic cannulation exhibited a better washout rate (99% vs. 57% of old blood was replaced in 12 s), lower blood stagnation volume (0.13 ml vs. 32.85 ml), and BRT (4.2 s vs. 7.1 s) than the RA cannulation in this simulated non-pulsatile case. Our findings suggest that RV diaphragmatic cannulation had a lower thrombosis risk and might be more favorable in a full RVAD-supported setting.
更多
查看译文
关键词
Heart failure,Right ventricular assist device,Blood flow dynamics,Thrombosis,Cannulation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要