Inter-device Reliability of the NPi-200 and NPi-300 Pupillometers

Journal of clinical neuroscience(2022)

引用 4|浏览5
暂无评分
摘要
The pupillary evaluation is an essential part of the neurological examination. Research suggests that the traditional examination of the pupil with a handheld flashlight has limited interrater reliability. Automated pupillometers were developed to provide an objective scoring of various pupillary parameters. The NPi-200 pupillometer is used for quantitative pupillary examinations, the NPi-300 was launched in July 2021 with enhanced features. The purpose of this study is to compare results from the NPi-200 to the NPi-300 to ensure that data are translatable across both platforms. This study examines the inter-device reliability of the NPi-200 compared to the NPi-300 in two cohorts: 20 patients at risk for cerebral edema and 50 healthy controls. Paired assessments of the devices were made from all participants. Each assessment included bilateral PLR readings within a 5-minute interval. Data showed high agreement between the two devices for the Neurological Pupil Index (NPi) reading (k = 0.94; CI: 0.91-0.99) and for pupil diameter assessment (k = 0.91; CI: 0.87-0.96). There is a very high level of agreement between the NPi-200 and NPi-300 among healthy controls and critically ill patients. Clinicians and researchers can interpret the results from either device equally.