This work reports the fabrication of a novel electrochemical sensor for the detection of the complex ion [Cu(CN)3](aq) 2? , one of the main components of the effluents from cyanide alkaline copper plating baths. This sensor was prepared by using a surface modified glassy carbon (GC) electrode with 1,8-dihydroxyantroquinone (1,8-DHAQ). It was characterized by electrochemical impedance, Raman spectroscopy and UV?Visible spectroscopy. The electrochemical detection of [Cu(CN)3](aq) 2? was performed by differential pulse voltammetry. The 1,8-DHAQ/GC electrochemical sensor exhibits good reproducibility and a linear range of 5.50 ? 10?7?3.81 ? 10?6 mol L?1, with a detection limit (LOD) of 1.20 ? 10?6 mol L?1, quantification limit (LOQ) of 3.97 ? 10?6 mol L?1, and robustness to interfering Cl?, NO3?, SO? ? y CO32? ions. Finally, a DFT approach suggests an almost parallel orientation of the bis(1,8dihydroxyanthraquinone) over a graphene domain of the glassy carbon surface, as well as the noncovalent nature of the interactions involved.
更多
查看译文
关键词
Electrochemical sensor,Glassy carbon,Adsorption,8-dihydroxyanthraquinone,[Cu(CN)3](aq),2?