Aim: Several devices are used for balloon tamponade. In Japan, metreurynters are widely used for cervical ripening; however, there is little evidence for their use in hemostatic uterine balloon tamponade. This study aimed to assess the non-inferiority of metreurynters as the balloon device for obstetric hemorrhage.Methods: Medical charts of all patients with obstetric hemorrhage of more than 1,000 ml who underwent balloon tamponade with metreurynters were retrospectively reviewed for five years at a tertiary referral university hospital.Results: A total of 89 uterine balloon tamponade cases were identified from medical records. Of these, 66 cases involved term postpartum hemorrhage (PPH) within 24 h after delivery (PPH group) and 23 involved other types of cases (other group), including cesarean scar pregnancy, cervical pregnancy, retained placenta, and others. In the PPH group, the average hemostasis rate was 93.9%, whereas the rate was 91.3% in the other group. Seven cases failed to achieve hemostasis with balloon tamponade only and required additional treatment. There were no adverse events related to balloon tamponade using metreurynters.Conclusions: Metreurynters used for balloon tamponade were non-inferior to Bakri balloons in hemostasis and complication rates, suggesting they are effective and appropriate for the management of obstetrical hemorrhage.