Aerosol Generation and Mitigation During Methacholine Bronchoprovocation Testing: Infection Control Implications in the Era of COVID-19

Respiratory care(2021)

引用 2|浏览11
暂无评分
摘要
BACKGROUND: Methacholine bronchoprovocation or challenge testing (MCT) is commonly performed to assess airway hyper-responsiveness in the setting of suspected asthma. Nebulization is an aerosol-generating procedure, but little is known about the risks of MCT in the context of the ongoing coronavirus disease 2019 (COV1D-19) pandemic. We aimed to quantify and characterize aerosol generMien during MCT by using different delivery methods and to assess the impact of adding a viral filter. METHODS: Seven healthy subjects performed simulated MCT in a near particle-free laboratory space with 4 different nehulizers and with a dosimeter. Two devices continuously sampled the ambient air during the procedure, which detected ultrafine particles, from 0.02-1 mu m, and particles of sizes 03, 0.3, 1.0, 2.0, 5.0, and 10 mu m, respectively. Particle generation was compared among all the devices, with and without viral filter placement. RESULTS: Ultrafine-particle generation during simulated MCT was significant across all the devices. Ultrafine-particle (0.02-1 mu m) concentrations decreased 77%-91% with the addition of a viral filter and varied significantly between unfiltered (P < .001) and filtered devices (P < .001). Ultrafine-particle generation was lowest when using the dosimeter with filtered Hudson nebulizer (1,258 +/- 1,644 particle/mL). Ultrafine-particle concentrations with the filtered nebulizer devices using a compressor were higher than particle concentrations detected when using the dosimeter: Monaghan (3,472 +/- 1,794 particles/mL), PARI (4,403 +/- 2,948), Hudson (6,320 +/- 1,787) and AirLife (9,523 +/- 5,098). CONCLUSIONS: The high particle concentrations generated during MCT pose significant infection control concerns during the COVID-19 pandemic. Particle generation during MCT was significantly reduced by using breath-actuated delivery and a viral filter, which offers an effective mitigation strategy.
更多
查看译文
关键词
Methacholine,Aerosol,Droplet,COVID-19,SARS-CoV-2,Infection Control,Asthma
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要