A 31-channel integrated "AC/DC" B-0 shim and radiofrequency receive array coil for improved 7T MRI

MAGNETIC RESONANCE IN MEDICINE(2022)

引用 13|浏览37
暂无评分
摘要
Purpose To test an integrated "AC/DC" array approach at 7T, where B-0 inhomogeneity poses an obstacle for functional imaging, diffusion-weighted MRI, MR spectroscopy, and other applications. Methods A close-fitting 7T 31-channel (31-ch) brain array was constructed and tested using combined Rx and Delta B-0 shim channels driven by a set of rapidly switchable current amplifiers. The coil was compared to a shape-matched 31-ch reference receive-only array for RF safety, signal-to-noise ratio (SNR), and inter-element noise correlation. We characterize the coil array's ability to provide global and dynamic (slice-optimized) shimming using Delta B-0 field maps and echo planar imaging (EPI) acquisitions. Results The SNR and average noise correlation were similar to the 31-ch reference array. Global and slice-optimized shimming provide 11% and 40% improvements respectively compared to baseline second-order spherical harmonic shimming. Birdcage transmit coil efficiency was similar for the reference and AC/DC array setups. Conclusion Adding Delta B-0 shim capability to a 31-ch 7T receive array can significantly boost 7T brain B-0 homogeneity without sacrificing the array's rdiofrequency performance, potentially improving ultra-high field neuroimaging applications that are vulnerable to off-resonance effects.
更多
查看译文
关键词
7 Tesla, echo planar imaging, functional MRI, In vivo B-0 shimming, integrated Delta B-0/Rx coil, multi-coil shimming, T-2* weighting