谷歌浏览器插件
订阅小程序
在清言上使用

The Experimental Study of a Russian Orthodontic Mini-Screw

Stomatologiâ(2021)

引用 0|浏览0
暂无评分
摘要
BACKGROUND:The aim of this research is the experimental study measuring stability of the orthodontic miniscrews «Turbo» designed in Russia in comparison with its foreign analogues, namely, «Vector Tas» (USA) and «BioRay» (Taiwan).MATERIAL AND METHODS:Four self-drilling orthodontic miniscrews of each manufacturer, i.e. «Vector Tas», USA, (10-mm length, 2-mm diameter), «BioRay», Taiwan, (10-mm length, 2-mm diameter), «Turbo», Russia, (9-mm length, 2-mm diameter), a total of 12 items, were inserted into native pig mandible sample. Their stability was estimated by torques using a dynamometer (Zahoransky AG, Germany) and «Periotest» device («Periotest M», Germany). This experiment was conducted in native pig mandible sample immediately after the screws' placement and in 7 days after loading at an angle 70°.RESULTS:After application of a load, the decreased torque values and increased Periotest values were registered in all orthodontic miniscrews.CONCLUSION:Orthodontic miniscrews «Turbo» designed in Russia are slightly inferior to «VectorTas» miniscrews and superior to «BioRay» miniscrews in primary stability and stability after 7 days under loading.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要