The Experimental Study of a Russian Orthodontic Mini-Screw
Stomatologiâ(2021)
摘要
BACKGROUND:The aim of this research is the experimental study measuring stability of the orthodontic miniscrews «Turbo» designed in Russia in comparison with its foreign analogues, namely, «Vector Tas» (USA) and «BioRay» (Taiwan).MATERIAL AND METHODS:Four self-drilling orthodontic miniscrews of each manufacturer, i.e. «Vector Tas», USA, (10-mm length, 2-mm diameter), «BioRay», Taiwan, (10-mm length, 2-mm diameter), «Turbo», Russia, (9-mm length, 2-mm diameter), a total of 12 items, were inserted into native pig mandible sample. Their stability was estimated by torques using a dynamometer (Zahoransky AG, Germany) and «Periotest» device («Periotest M», Germany). This experiment was conducted in native pig mandible sample immediately after the screws' placement and in 7 days after loading at an angle 70°.RESULTS:After application of a load, the decreased torque values and increased Periotest values were registered in all orthodontic miniscrews.CONCLUSION:Orthodontic miniscrews «Turbo» designed in Russia are slightly inferior to «VectorTas» miniscrews and superior to «BioRay» miniscrews in primary stability and stability after 7 days under loading.