Background and purpose: Aperture-based complexity metrics have been suggested as a method to score complexity of volumetric modulated arc therapy (VMAT) plans. The purpose of this study was to evaluate the edge area metric (EAM) for clinical VMAT plans on a control point and treatment plan level. Materials and methods: EAM on a control point level was evaluated based on film measurements of 18 static beam openings originating from VMAT plans. EAM on a treatment plan level (arithmetic mean value of EAM scores for control points) was evaluated based on measurements with the Delta(4)(R) for 200 VMAT plans for four different treatment sites: pelvic, thorax, head and neck, and prostate. Measurements were compared to calculations and dose difference and gamma pass rates were evaluated. Results: EAM scores on a control point level correlated with Pearson's r-values of -0.96 and -0.77 to dose difference and gamma pass rates, respectively. The prostate plans had the highest average EAM score. A connection between smaller PTVs and higher EAM scores was found. No correlation between the evaluation result and EAM on a plan level was found. Conclusions: EAM on a control point level was shown to correlate to the difference between measured and calculated 2D dose distributions of clinical VMAT beam openings. No correlation was found for EAM on a plan level for clinical treatment plans.