Measurement of the Human Cadaver Ossicle Vibration Amplitude by Fiber-Optic Interferometry
2020 43RD INTERNATIONAL CONVENTION ON INFORMATION, COMMUNICATION AND ELECTRONIC TECHNOLOGY (MIPRO 2020)(2020)
摘要
In this paper we present the results obtained by contactless measurement of vibration of the incus inside the middle ear of a human cadaver, exposed to an external calibrated sound source. The measurement configuration was a Michelson fiber-optic interferometer based on a 3x3 fiber optic coupler, with a VCSEL source. The interferometric signal fading was solved using two quasiquadrature signals and specialized signal processing. The ossicle reflection was augmented by a very light piece of retroreflective material attached to the bone. We measured the amplitude of vibration at different sound pressure levels, independently monitored by a decibel meter placed in the proximity of the cadaver ear. The accuracy of the measurement technique was proved by using a calibrated piezo transducer exposed to vibrate with amplitude of 7nm at 1kHz. We performed the experiments on 8 cadavers in total. The measured amplitudes were found to be in the range between 1 pm and 1 nm, under the sound excitation of 40-90 dB SPL @ 1kHz, respectively.
更多查看译文
关键词
fiber-optic sensors,vibration measurement,implantable microphone,hearing aid
AI 理解论文
溯源树