Dosimetric and Geometric End-to-end Accuracy of a Magnetic Resonance Guided Linear Accelerator

Physics and imaging in radiation oncology(2020)

引用 0|浏览11
暂无评分
摘要
The introduction of real-time imaging by magnetic resonance guided linear accelerators (MR-Linacs) enabled adaptive treatments and gating on the tumor position. Different end-to-end tests monitored the accuracy of our MR-Linac during the first year of clinical operation. We report on the stability of these tests covering a static, adaptive and gating workflow. Film measurements showed gamma passing rates of 96.4% ± 3.4% for the static tests (five measurements) and for the two adaptive tests 98.9% and 99.99%, respectively (criterion 2%/2mm). The gated point dose measurements in the breathing phantom were 2.7% lower than in the static phantom.
更多
查看译文
关键词
MR-Linac,Treatment adaption,Gating,End-to-end test