[Application of Hemodynamic Optimization in the Design of Artificial Heart].

PubMed(2020)

引用 2|浏览18
暂无评分
摘要
Heart failure is one kind of cardiovascular disease with high risk and high incidence. As an effective treatment of heart failure, artificial heart is gradually used in clinical treatment. Blood compatibility is an important parameter or index of artificial heart, and how to evaluate it through hemodynamic design and in vitro hemolysis test is a research hotspot in the industry. This paper first reviews the research progress in hemodynamic optimization and in vitro hemolysis evaluation of artificial heart, and then introduces the research achievements and progress of the team in related fields. The hemodynamic performance of the blood pump optimized in this paper can meet the needs of use. The normalized index of hemolysis obtained by in standard vitro hemolysis test is less than 0.1 g/100 L, which has good hemolysis performance in vitro. The optimization method described in this paper is suitable for most of the development of blood pump and can provide reference for related research work.