Monte Carlo Optical Simulations of a Small FoV Gamma Camera. Effect of Scintillator Thicknesses and Septa Materials

Crystals(2019)

Cited 8|Views10
No score
Abstract
Optical Monte Carlo simulations have been extensively used for the accurate modeling of light transport in scintillators for the improvement of detector designs. In the present work, a GATE Monte Carlo toolkit was used to study the effect of scintillator thicknesses and septa materials in the performance parameters evaluation of a commercially available small animal gamma-optical camera, named “γ-eye”. Firstly, the simulated γ-eye system was validated against experimental data. Then, part of the validated camera was modeled defining all of the optical properties by means of the UNIFIED model of GATE. Different CsI:Na scintillator crystals with varying thicknesses (from 4 mm up to 6 mm) and different reflector (septa) materials were simulated and compared in terms of sensitivity, light output and spatial resolution. Results have demonstrated the reliability of the model and indicate that the thicker crystal array presents higher sensitivity values, but degraded spatial resolution properties. Moreover, the use of black tape around crystals leads to an improvement in spatial resolution values compared to a standard white reflector material.
More
Translated text
Key words
GATE,Monte Carlo simulation,light transport,optical simulation,scintillator crystals