Use of CAD-based pre-bent implants reduces theatre time in orbital floor reconstruction: results of a prospective study.

BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY(2020)

引用 7|浏览14
暂无评分
摘要
In this prospective study we evaluated the duration of insertion and the accuracy of fitting of computer-aided design (CAD)-based pre bent orbital floor plates compared with non-preformed orbital plates in reconstruction of the orbital floor. Thirty-six patients with unilateral fractures of the orbital floor were included; pre-bent plates were used in 25 and non-preformed plates in 11. Preoperative computed tomography (CT) scans were used for CAD of individualised implants. The anatomy of the affected orbit was virtually reconstructed "slice by slice". Individually pre-bent plates were generated using a stereolithographic model of the reconstructed orbit. The mean (SD) duration of insertion was significantly reduced when pre-bent orbital plates were used (5.5 (5.4) min) compared with non-preformed meshes (11.1 (7.7) min). The congruence of pre-bent plates to the infraorbital rim did not differ from that of non-preformed plates. The accuracy of fit was rated as "accurate" in 24 cases. CAD-based individualised titanium meshes reduce theatre time compared with non-preformed orbital plates. Our results confirm the efficacy of CAD-based pre-bent plates in reconstruction of the orbital floor. (C) 2019 The British Association of Oral and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.
更多
查看译文
关键词
intraoperative time comparison,individually pre-bent titanium mesh,computer-aided design,orbital floor fracture,orbital floor reconstruction