Spatial Resolution Of A Spherical X-Ray Crystal Spectrometer At Various Magnifications

REVIEW OF SCIENTIFIC INSTRUMENTS(2016)

Cited 0|Views7
No score
Abstract
A high spatial resolution of a few mu m is often required for probing small-scale high-energy-density plasmas using high resolution x-ray imaging spectroscopy. This resolution can be achieved by adjusting system magnification to overcome the inherent limitation of the detector pixel size. Laboratory experiments on investigating the relation between spatial resolution and system magnification for a spherical crystal spectrometer are presented. Tungsten L beta(2) rays from a tungsten-target micro-focus x-ray tube were diffracted by a Ge 440 crystal, which was spherically bent to a radius of 223 mm, and imaged onto an x-ray CCD with 13-mu m pixel size. The source-to-crystal (p) and crystal-to-detector (q) distances were varied to produce spatial magnifications (M = q/p) ranging from 2 to 10. The inferred instrumental spatial width reduces with increasing system magnification M. However, the experimental measurement at each M is larger than the theoretical value of pixel size divided by M. Future work will focus on investigating possible broadening mechanisms that limit the spatial resolution. Published by AIP Publishing.
More
Translated text
Key words
spectrometer,spatial resolution,crystal,x-ray