Studying the Counting Rate Capability of A Glass Multigap Resistive Plate Chamber at an Increased Operating Temperature

Instruments and experimental techniques(2013)

引用 2|浏览41
暂无评分
摘要
It is shown that a multigap resistive plate chamber made of commercial float glass is capable of sustaining high counting rates at an increased operating temperature. Two glass chambers were investigated on the test beamline of the U-70 accelerator at the Institute for High Energy Physics. The required radiation flux density at the detector was produced by means of radioactive sources. The time resolution of 80 ps or better was attained at a rate of ∼20 kHz/cm 2 and an operating temperature of 45°C.
更多
查看译文
关键词
Rate Capability,Radioactive Source,Volume Resistivity,High Counting Rate,High Voltage Electrode