谷歌浏览器插件
订阅小程序
在清言上使用

Mass Spectrometric Calibration of Controlled Fluoroform Leak Rate Devices: Technique and Uncertainty Analysis

crossref(2003)

引用 23|浏览3
暂无评分
摘要
Controlled leak rate devices of fluoroform on the order of 10{sup -8} atm {center_dot} cc sec{sup -1} at 25 C are used to calibrate QC-1 War Reserve neutron tube exhaust stations for leak detection sensitivity. Close-out calibration of these tritium-contaminated devices is provided by the Gas Dynamics and Mass Spectrometry Laboratory, Organization 14406, which is a tritium analytical facility. The mass spectrometric technique used for the measurement is discussed, as is the first principals calculation (pressure, volume, temperature and time). The uncertainty of the measurement is largely driven by contributing factors in the determination of P, V and T. The expanded uncertainty of the leak rate measurement is shown to be 4.42%, with a coverage factor of 3 (k=3).
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要