Calibration of a broadband acoustic transducer with a standard spherical target in the near field.

JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA(2015)

引用 7|浏览1
暂无评分
摘要
This paper investigates the applicability of calibrating a broadband acoustic system in the near field. The calibration was performed on a single transducer with a mono-static configuration using a single standard target, a 25-mm tungsten carbide sphere in the nearfield of both the transducer and the sphere. A theoretical model was developed to quantify the nearfield effect. Numerical simulations revealed that the frequency responses at different distances varied significantly, the null positions were essentially invariant-a unique characteristic for determination of the compressional and shear wave speeds in the calibration sphere. The calibration curves obtained in the near field could be applied to farfield once the nearfield effects were accounted for. Since the transducer was located in the near field, the signal-to-noise ratio was high, resulting in a much wider useable bandwidth than the nominal bandwidth. The resultant calibration uncertainty, i.e., root-mean-square uncertainty over the entire usable frequency band was 1.05 dB and reduces to 0.33 dB when the regions corresponding to nulls were excluded. The methods reported here could potentially be applied to the calibration of multibeam and broadband echosounder/sonar systems since it is difficult to meet the farfield condition for outermost beams when shipboard calibrations are needed.
更多
查看译文
关键词
broadband acoustic transducer,standard spherical target