WE-C-M100F-09: Dosimetric Comparison of Linac-IMRT and Helical Tomotherapy (HT) for Head and Neck Cancer

MEDICAL PHYSICS(2007)

引用 0|浏览7
暂无评分
摘要
Purpose: To perform a dosimetric comparison between Linac and Helical Tomotherapy‐based IMRT on head and neck (HN Rc = the ratio of the PTV coverage to PTV volume, where the former was defined as the volume enclosed by the 66 Gy isodose surface; Biologically equivalent doses (BED) to organs at risk (OAR) and PTV dose homogeneity were also studied. The tolerance range or TR (standard deviation/PTV average dose) was used as a surrogate for PTV dose homogeneity evaluation. Results: R95 results indicated that both IMRT techniques produced comparable conformal plans. Rc values showed that HT plans generally provided better tumor coverage. TR results suggest that PTV dose homogeneity was better for HT plans. Finally, the average OAR BEDs showed a trend of better normal tissue sparing with the HT plans. The exception was for the spinal cord, in which the maximum BED using HT was slightly lower than the maximum BED using linac‐IMRT. Conclusion: This study suggests that HT plans had in general better dosimetric characteristics, especially regarding tumor coverage, PTV dose homogeneity and normal tissue sparing in physician approved plans. Dose reductions to OAR may not yield any clinical differences in outcome in virtue of the delivered OAR doses which are well below normal tissue tolerance.
更多
查看译文
关键词
helical tomotherapy,neck cancer,dosimetric comparison,linac-imrt