On-Chip Immunoassay for Determination of Urinary Albumin

Sensors(2009)

引用 27|浏览6
暂无评分
摘要
An immunoassay performed on a portable microfluidic device was evaluated for the determination of urinary albumin. An increase in absorbance at 500 nm resulting from immunoagglutination was monitored directly on the poly(dimethylsiloxane) (PDMS) microchip using a portable miniature fibre-optic spectrometer. A calibration curve was linear up to 10 mg L-1 (r(2) = 0.993), with a detection limit of 0.81 mg L-1 (S/N = 3). The proposed system showed good precision, with relative standard deviations (RSDs) of 5.1%, when evaluated with 10 mg L-1 albumin (n = 10). Determination of urinary albumin with the proposed system gave results highly similar to those determined by the conventional spectrophotometric method using immunoturbidimetric detection (r(2) = 0.995; n = 15).
更多
查看译文
关键词
microalbuminuria,microfluidic system,on-chip immunoassay,poly(dimethylsiloxane)