Landsat-5 TM Reflective-Band Absolute Radiometric Calibration.

IEEE transactions on geoscience and remote sensing(2004)

引用 97|浏览22
暂无评分
摘要
The Landsat-5 Thematic Mapper (TM) sensor provides the longest running continuous dataset of moderate spatial resolution remote sensing imagery, dating back to its launch in March 1984. Historically, the radiometric calibration procedure for this imagery used the instrument's response to the Internal Calibrator (IC) on a scene-by-scene basis to determine the gain and offset of each detector. Due to observed degradations in the IC, a new procedure was implemented for U.S.-processed data in May 2003. This new calibration procedure is based on a lifetime radiometric calibration model for the instrument's reflective bands (1-5 and 7) and is derived, in part, from the IC response without the related degradation effects and is tied to the cross calibration with the Landsat-7 Enhanced Thematic Mapper Plus. Reflective-band absolute radiometric accuracy of the instrument tends. to be on the order of 7% to 10%, based on a variety of calibration methods.
更多
查看译文
关键词
absolute calibration,characterization,Internal Calibrator (IC),Landsat,Landsat-5 (L5) Thematic Mapper (TM),Landgat-7 (L7) Enhanced Thematic Mapper Plus (ETM plus ),lookup table (LUT),radiance,radiometry,reflectance,relative spectral response,vicarious