谷歌浏览器插件
订阅小程序
在清言上使用

Parametric Evaluation For Programmable Temperature Vaporisation Large Volume Injection In Gas Chromatographic Determination Of Polybrominated Diphenyl Ethers

INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY(2010)

引用 14|浏览16
暂无评分
摘要
This work is a thorough investigation on the major operating parameters of the programmable temperature vaporisation (PTV) inlet used for gas chromatographic injection, including injection mode and volume, inlet temperature, vent and purge flow rates. The results clearly demonstrate the advantage of large volume injection in enhancing the detection of polybrominated diphenyl ethers (PBDEs). Partial loss of injected PBDEs occurred during solvent venting and due to incomplete sample transfer. Such loss was minimised by lowering the initial inlet temperature and vent flow and elevating the final inlet temperature. The results show that 50 mL/min vent flow, as low as 0 degrees C initial and higher than 300 degrees C final inlet temperatures produced the relatively high responses. Two mass spectrometric parameters were also evaluated. Indoor dust, lake sediment and human placenta tissue samples were analysed to demonstrate reliability and sensitivity improvement of the PTV large volume injection.
更多
查看译文
关键词
gas chromatography, PTV, large volume injection, polybrominated diphenyl ethers
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要