Analysis of part-per-billion level of arsine and phosphine in light hydrocarbons by capillary flow technology and dielectric barrier discharge detector.

Journal of Chromatography A(2010)

引用 17|浏览5
暂无评分
摘要
A practical gas chromatographic procedure has been developed and implemented for the measurement of arsine and phosphine in hydrocarbons such as propylene at the part-per-billion level. The successful measurement of arsine and phosphine at the level mentioned was attained by incorporating a large volume injection technique to increase the mass of solutes delivered for sensitivity improvement, capillary flow technology to keep the matrix from entering the detector by either back-flushing through the inlet vent, or by heart-cutting if required, and dielectric barrier discharge detector operating in argon mode for sensitivity enhancement, as well as offering improved selectivity towards the solutes cited. Using the technique described a complete analysis can be conducted in less than 4min. A relative standard precision of less than 1.7% was achieved with repeated injections at the concentration level of 25 and 125ppb (v/v) each of arsine and phosphine in nitrogen with a practical detection limit at the 5ppb (v/v) level. Correlation coefficients of greater than 0.9999 were obtained for arsine and phosphine over a range from 10 to 2500ppb (v/v). The analytical methodology was proven to be reliable in continuous operation during the first 6 months of deployment.
更多
查看译文
关键词
Gas chromatography,Capillary flow technology,Argon dielectric barrier discharge detector,Arsine,Phosphine,LVGIS