Complete Prevention of Bubbles in a PDMS-Based Digital PCR Chip with a Multifunction Cavity

BIOSENSORS-BASEL(2024)

引用 0|浏览18
暂无评分
摘要
In a chamber-based digital PCR (dPCR) chip fabricated with polydimethylsiloxane (PDMS), bubble generation in the chambers at high temperatures is a critical issue. Here, we found that the main reason for bubble formation in PDMS chips is the too-high saturated vapor pressure of water at an elevated temperature. The bubbles should be completely prevented by reducing the initial pressure of the system to under 13.6 kPa to eliminate the effects of increased-pressure water vapor. Then, a cavity was designed and fabricated above the PCR reaction layer, and Parylene C was used as a shell covering the chip. The cavity was used for the negative generator in sample loading, PDMS degassing, PCR solution degassing in the digitization process and water storage in the thermal reaction process. The analysis was confirmed and finally achieved a desirable bubble-free, fast-digitization, valve-free and no-tubing connection dPCR.
更多
查看译文
关键词
digital PCR chip,polydimethylsiloxane,multifunction cavity,water loss,bubble generation,parylene C shell,negative pressure environment,saturated vapor pressure of water