System Accuracy and Interference Evaluation of a New Glucose Dehydrogenase-Based Blood Glucose Meter for Patient Self-Testing.

Andreas Pfützner, Daiva Kalasauske,Mina Hanna,Daniela Sachsenheimer, Gerhard Raab, Silvia Weissenbacher,Nicole Thomé

Journal of diabetes science and technology(2023)

引用 0|浏览5
暂无评分
摘要
New European medical device regulations require the performance of postmarketing surveillance evaluations for blood glucose meters (BGMs). We conducted an ISO15197:2015-conform system performance evaluation with the approved glucose dehydrogenase (GDH)-based Wellion NEWTON BGM. One hundred subjects were enrolled into the study (44 female, 56 male, 43 healthy subjects, 23 type 1 diabetes, 34 type 2 diabetes, age: 53.7 ± 15.8 years). In addition, manipulated heparinized whole blood was used for a laboratory interference test with ten selected substances (interference definition: substance-induced bias > 10%). The mean absolute relative difference (MARD) was 4.7%, and 100% of the values were in zones A (99.7%) and B (0.3%), respectively, of the consensus error grid. Interference was observed with xylose only, which is a known interfering substance for GDH-based BGMs.
更多
查看译文
关键词
blood glucose meter,system accuracy,dehydrogenase-based,self-testing