谷歌浏览器插件
订阅小程序
在清言上使用

Enhancement of Real-Time Resonance Tracking in Electrothermally Actuated Cantilever Sensor with Optimized Phase Characteristic

Journal of physics Conference series(2019)

引用 0|浏览9
暂无评分
摘要
Non-ideal phase responses on electro-thermally actuated piezoresistive cantilever sensors have led the phase-locked loop (PLL) systems into difficulties for real-time sensing applications. These outcomes are caused by thermal-parasitic coupling from the actuating part to the sensing part. Minimizing or eliminating parasitic effects is necessary to obtain an optimized phase response. To realize this, we adjusted the voltage supply of the sensing part, which is in form of a full Wheatstone bridge (WB). By increasing the WB supply voltage (V WB), the phase response can be enhanced. Alternatively, a reference signal that differentially eliminates the parasitic parameter from the sensor output was employed. To investigate the resulting optimized phase response under real-time measurement conditions, two different microcantilevers were connected to an MFLI lock-in amplifier + PLL system (Zurich Instruments). Measurement results exhibited a good sensor performance under varying humidity and temperature conditions.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要