The Fatal Consequence Of Inappropriate Therapy In A Single Vf Zone Primary Prevention Defibrillator

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(2021)

引用 0|浏览13
暂无评分
摘要
We report a case of inappropriate implantable cardioverter defibrillator (ICD) therapy due to single ventricular fibrillation (VF) zone programming leading to patient death. A remote transmission was received from a patient with a cardiac resynchronization therapy-defibrillator showing sinus tachycardia in the VF detection zone initiating inappropriate shocks and resulting in shock refractory VF. This case report highlights the importance of manufacturer specific ICD programming. In devices without discrimination in the VF zone, a higher rate single VF detection zone and/or addition of a ventricular tachycardia zone with supraventricular tachycardia discrimination should be considered.
更多
查看译文
关键词
CRT&#8208, D, inappropriate therapy, primary prevention, proarrhythmia
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要