谷歌浏览器插件
订阅小程序
在清言上使用

Mitigating Hypovolemia-Induced Miscalibration of Photoplethysmogram-Derived Blood Pressure

42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20(2020)

引用 1|浏览29
暂无评分
摘要
Pulse transit time (PTT) is a hemodynamic indicator that may be obtained non-invasively using photoplethysmogram (PPG) signals for continuous blood pressure (BP) monitoring. Among the most promising applications of this technology are military and civilian trauma cases, where reduced blood volume due to hemorrhage, or absolute hypovolemia, is the leading preventable cause of death. However, the drawback of this method is that it requires calibration for each patient; additionally, changes in physiological state may affect PTT calibration. In this work, a porcine model (n = 6) was used to demonstrate that changes in blood volume lead to miscalibration of PTT for BP estimation. To mitigate hypovolemia-induced miscalibration, this work first defines a template-based signal quality index (SQI) for characterizing the morphology of PPG signals; it is then shown that the subject-specific calibration of SQI to BP is more robust to changes in blood volume than PTT. Though changes in PPG signal quality are not necessarily specific to changes in BP, these results suggest that PPG-based monitoring systems may benefit from incorporating morphological information for cuffless BP estimation in trauma settings.
更多
查看译文
关键词
Animals,Blood Pressure,Blood Pressure Determination,Humans,Hypovolemia,Photoplethysmography,Pulse Wave Analysis,Swine
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要