Systematic review of microvascular coupling devices for arterial anastomoses in free tissue transfer.
Laryngoscope investigative otolaryngology(2020)
Abstract
BACKGROUND:Coupling devices have become commonplace in performing venous anastomoses during microvascular free tissue transfer (FTT). However, arterial anastomoses are still most commonly performed using traditional microvascular suture techniques.
OBJECTIVE:To describe the safety and feasibility of using microvascular coupling devices for free flap arterial anastomosis.
METHODS:A systematic review of English language literature was performed for studies that investigated the use of an arterial coupler for microvascular FTT in human patients. A comprehensive search of MEDLINE (January 1948 to August 2018), EMBASE (January 1974 to August 2018), and Web of Science was performed.
RESULTS:Fifteen studies were included. All studies were retrospective case series. A combined total of 395 arterial anastomoses were attempted with a coupling device. All studies except for one used the 3M Unilink/Synovis coupling device. One study used a novel absorbable coupling device. The coupling device was aborted and converted to a suture technique in 8.4% of attempted anastomoses. Rupture of the anastomotic device was reported in only 1 patient (0.3%). Thrombosis was also infrequent at 1.9%. The quality assessment showed a high risk of bias in all studies.
CONCLUSION:In selected patients, coupling devices for arterial anastomoses have a good success rate with low rates of thrombosis based on limited quality evidence.