Novel Histopathologic And Immunohistochemical Observations In Explanted Orbital Peri-Implant Capsules

CURRENT EYE RESEARCH(2021)

引用 1|浏览3
暂无评分
摘要
PurposeTo describe the histopathologic and immunohistochemical characteristics of peri-orbital implant capsule and correlate the observed changes with delayed implant extrusion.Materials and methodsProspective, ex-vivo, histopathologic, immunohistochemical study of peri-implant capsules excised from capsules surrounding a poly-methyl metha acrylate (PMMA) implant. Thirteen capsules were harvested and divided into two groups. Group 1 (implant extrusion group) consisted of capsules harvested from around exposed/extruded implants and Group 2 (implant non-extrusion group) from implants that were surgically exchanged for volume augmentation. Data collected included demography, clinical presentations, etiology of the explantation, age of the capsule in months, inflammatory cells noted/high power field (HPF), CD3, CD20, and CD68, percentage Masson's trichrome staining and mean capsular thickness.ResultsMean patient age at presentation was 33 +/- 13 years. Of the 13 capsules included, 7 belonged to group 1and 6 to group 2. Mean age of the fibro-collagenous membranes were 33 +/- 57 months (median 13 months).Median CD3, CD20, and CD68 positivity was 80%, 9%, and 7% denoting a predominant T-lymphocytic response. Mean capsular thickness was 733 +/- 422 mu (median 678 mu). Age of the harvested capsule showed a linear correlation with inflammatory cells/HPF (r = 0.93, p < .0001, 95% C.I. 0.73 to 0.98), with CD3 positivity (r = 0.6, p = .04, 95% C.I. 0.02 to 0.9), with capillary proliferation/HPF (r = 0.85, p = .003, 95% C.I. 0.43 to 0.96) and with percentage of cells showing Masson's trichrome stain positivity (r = 0.71, p = .03, 95% C.I. 0.08 to 0.93). In the extrusion group, the capsules demonstrated high inflammation and low fibrosis.ConclusionsPMMA orbital spherical implant incites a predominantly T-cell mediated inflammatory response which clinically presents as a peri-implant capsule. Increasing inflammation in the absence of significant fibrosis may be responsible for delayed PMMA implant exposure and extrusion.
更多
查看译文
关键词
Implant extrusion, orbital, complication, PMMA
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要